| Availability: | |
|---|---|
XRME20220301033
XINDRAY
Specifications:
| Model name | CNME030804:pH,pCO2,pO2,k+,Na+,Cl+,Ca++,Hct CNME030804A:pH,pCO2,pO2,Hct CNME030804B:pH,pCO2,pO2,k+,Na+,Cl+,HcT CNME030804C:pH,pCO2,pO2,k+,Na+,Cl+,Ca++,Hct,AutoQC CNME030804D:pH,pCO2,pO2,Hct,AutoQC CNME030804E:pH,pCO2,pO2,k+,Na+,Cl+,Hct,AutoQC |
| Analysis Time | <90s |
| Sample Volume | 95uL Whole blood /50uL Capillary blood |
| Calibration | Automatic, programmable 1-point and 2-point calibration |
| Data Storage | >10000 |
| Display | 10.4" TFT touch screen |
| Battery | Nickel-metal hydride battery,3.0Ah |
| Printer | Build-in Thermal printer,70mm |
| Barcode Reader | Up to 14 kinds barcode type |
| Sample | Whole blood, Serum, Plasma, Dialysate, CSF |
| Instrument | 400(W)*574(H)*344(D)mm3 |
| CAL PAK | 258(W)*180(H)*70(D)mm2 |
| AUTOQC | 103(W)*180(H)*70(D)mm2 |
| Weight | 18kg |
| Power supply | 100-240V~50/60Hz |
| pH(TC) | pH temperature corrected |
| pCO2(TC) | pCO2 temperature corrected |
| pO2(TC) | pO2 temperature corrected |
| TCO2 | Total carbon dioxide |
| HCO3- | Actual plasma bicarbonate |
| BEb | Base Excess in blood |
| BEecf | Base Excess in extra cellular fluids |
| SBC | Standard bicarbonate |
| sO2% | Calculated oxygen saturation percentage |
| Ca++(7.4 | lonized calcium "normalized" to pH 7.4 |
| TCa | Total calcium content |
| AG | Anion gap |
| RI | Respiratory Index |
| AaDO2 | Alveolar artillery oxygen gradient |
| O2SAT | Oxygen Saturation |
| tHb(c) | Hemoglobin(Calculated) |
Specifications:
| Model name | CNME030804:pH,pCO2,pO2,k+,Na+,Cl+,Ca++,Hct CNME030804A:pH,pCO2,pO2,Hct CNME030804B:pH,pCO2,pO2,k+,Na+,Cl+,HcT CNME030804C:pH,pCO2,pO2,k+,Na+,Cl+,Ca++,Hct,AutoQC CNME030804D:pH,pCO2,pO2,Hct,AutoQC CNME030804E:pH,pCO2,pO2,k+,Na+,Cl+,Hct,AutoQC |
| Analysis Time | <90s |
| Sample Volume | 95uL Whole blood /50uL Capillary blood |
| Calibration | Automatic, programmable 1-point and 2-point calibration |
| Data Storage | >10000 |
| Display | 10.4" TFT touch screen |
| Battery | Nickel-metal hydride battery,3.0Ah |
| Printer | Build-in Thermal printer,70mm |
| Barcode Reader | Up to 14 kinds barcode type |
| Sample | Whole blood, Serum, Plasma, Dialysate, CSF |
| Instrument | 400(W)*574(H)*344(D)mm3 |
| CAL PAK | 258(W)*180(H)*70(D)mm2 |
| AUTOQC | 103(W)*180(H)*70(D)mm2 |
| Weight | 18kg |
| Power supply | 100-240V~50/60Hz |
| pH(TC) | pH temperature corrected |
| pCO2(TC) | pCO2 temperature corrected |
| pO2(TC) | pO2 temperature corrected |
| TCO2 | Total carbon dioxide |
| HCO3- | Actual plasma bicarbonate |
| BEb | Base Excess in blood |
| BEecf | Base Excess in extra cellular fluids |
| SBC | Standard bicarbonate |
| sO2% | Calculated oxygen saturation percentage |
| Ca++(7.4 | lonized calcium "normalized" to pH 7.4 |
| TCa | Total calcium content |
| AG | Anion gap |
| RI | Respiratory Index |
| AaDO2 | Alveolar artillery oxygen gradient |
| O2SAT | Oxygen Saturation |
| tHb(c) | Hemoglobin(Calculated) |