Availability: | |
---|---|
XRME210210132
XINDRAY
1) In-built battery backup lo support 30 minutes after power off
2) Disinfection: Rinse, hot disinfection and chemic disinfection. Circulated and closed disinfection system
3) Automatically turn off machine once disinfection process completes
4) Diameter of blood pump can be adjusted to be flexible for adult and child
5) Be applicable for various concentration ratio formula
6) Store treatments information in form of curve profile
7) Full function digital selftest process
8) With visual and auditive alarm. character message shows alarm item and troubleshooting on screen
1) Air bubble monitor: optical and ultrasonic double control, can monitor blood level and accumulate air bubble
2) With safe and sensitive blood leakage monitor device
3) Complete water-way and electric way apart design, individual waterway control system to ensure the security
4) Dialysate temperature with high precision, safety control at 35-39℃
5) Protection from intercross infection, fluid supply loop and return loop is completely divided during disinfection to
6) Avoid pollution of whole machine in the case of mis-sucking disinfection liquid
7) Conductivity control and protection: conductivity A. mix. conductivity, water outlet conductivity, insure patient's security 8) Adjustable Na+ curve profile, with multiplicate optional auto-adjust program, can realize individual dialysis and ensure patient's safety
Technical Specification:
Dimension | 1450*550*460 |
Net weight | Aprox.60kg |
Power supply | AC220V, -20%~+10%, 50~60HZ |
Water supply | Pressure range: 0~8.0bar |
Temperature range: 0~35℃ | |
Water flow: ≥0.3L/min | |
VP | -700~+800mmhg, Accuracy±5mmhg |
AP | -700~+800mmhg, Accuracy±5mmhg |
TMP | -700~+800mmhg, Accuracy±5mmhg |
Blood Flow | 0~700ml/Min |
Dialysate Flow | 0~300~500~800~1000ml/Min(adjustable) |
Dialysate Temperature | 30~41℃, Accuracy ±0.5℃ |
Dialysate Conductivity | 0~25 MS/cm, Accuracy ±0.1ms/cm |
UF control | UF rate: 0~6000ml/h |
Accuracy: ±30ml/h or ±1% | |
Substitution Fluid Flow | 0~39L/h(HDF-Online) |
0~6L/h(HDF-bibag) | |
Heparin injection | 0~10ml/H, Accuracy ±0.1ml/H |
Bolus: 0~10ml/time | |
Syringe size: 20ml, 30ml, 50ml | |
Air Detector | Detecting≥0.02ML/single bubble |
Detect the bubbles | |
Blood leak monitor | Detecting blood leak≥0.35ml/min(HCT 32%) |
1) In-built battery backup lo support 30 minutes after power off
2) Disinfection: Rinse, hot disinfection and chemic disinfection. Circulated and closed disinfection system
3) Automatically turn off machine once disinfection process completes
4) Diameter of blood pump can be adjusted to be flexible for adult and child
5) Be applicable for various concentration ratio formula
6) Store treatments information in form of curve profile
7) Full function digital selftest process
8) With visual and auditive alarm. character message shows alarm item and troubleshooting on screen
1) Air bubble monitor: optical and ultrasonic double control, can monitor blood level and accumulate air bubble
2) With safe and sensitive blood leakage monitor device
3) Complete water-way and electric way apart design, individual waterway control system to ensure the security
4) Dialysate temperature with high precision, safety control at 35-39℃
5) Protection from intercross infection, fluid supply loop and return loop is completely divided during disinfection to
6) Avoid pollution of whole machine in the case of mis-sucking disinfection liquid
7) Conductivity control and protection: conductivity A. mix. conductivity, water outlet conductivity, insure patient's security 8) Adjustable Na+ curve profile, with multiplicate optional auto-adjust program, can realize individual dialysis and ensure patient's safety
Technical Specification:
Dimension | 1450*550*460 |
Net weight | Aprox.60kg |
Power supply | AC220V, -20%~+10%, 50~60HZ |
Water supply | Pressure range: 0~8.0bar |
Temperature range: 0~35℃ | |
Water flow: ≥0.3L/min | |
VP | -700~+800mmhg, Accuracy±5mmhg |
AP | -700~+800mmhg, Accuracy±5mmhg |
TMP | -700~+800mmhg, Accuracy±5mmhg |
Blood Flow | 0~700ml/Min |
Dialysate Flow | 0~300~500~800~1000ml/Min(adjustable) |
Dialysate Temperature | 30~41℃, Accuracy ±0.5℃ |
Dialysate Conductivity | 0~25 MS/cm, Accuracy ±0.1ms/cm |
UF control | UF rate: 0~6000ml/h |
Accuracy: ±30ml/h or ±1% | |
Substitution Fluid Flow | 0~39L/h(HDF-Online) |
0~6L/h(HDF-bibag) | |
Heparin injection | 0~10ml/H, Accuracy ±0.1ml/H |
Bolus: 0~10ml/time | |
Syringe size: 20ml, 30ml, 50ml | |
Air Detector | Detecting≥0.02ML/single bubble |
Detect the bubbles | |
Blood leak monitor | Detecting blood leak≥0.35ml/min(HCT 32%) |